CRT: |
Type : |
Rectangular with integral graticule. |
Display Area : |
8 x 10 div ( 1 div = 1 cm ) |
Accelerating Voltage : |
12kV. |
Phosphor : |
P31. |
VERTICAL AMPLIFIERS ( CH 1 and
CH 2 ) |
Sensitivity : |
5 mV / div to 5 V / div ( X1 mode ). |
1 mV / div to 1 V / div ( X5 mode ). |
1-2-5 sequence, 10 step attenuator. Vernier
control provides fully adjustable attenuation between
steps. |
Accuracy : |
± 3 % ( X1 mode );± 5 % ( X5 mode ). |
Input Resistance: |
1 MΩ ± 2 %. |
Input Capacitance: |
25 pF ± 10 pF. |
Frequency Response: |
5 mV to 5 V / div ( CAL ):DC to 60 MHz (-3
dB).
1
mV / div (CAL):DC to 20 MHz (-3 dB). |
Rise Time: |
Approximately 5.8 ns. |
Operaing Modes: |
CH1: |
CH 1, single trace. |
CH2: |
CH 2, single trace. |
ALT: |
Dual trace, alternating. |
CHOP: |
dual trace, chopped. |
ADD: |
algebraic sum of CH 1 + CH 2. |
Polarity Reeversal (Invert): |
CH 2 only. |
Maximum Input Voltae: |
400 V dc + ac peak. |
Maximum Undistorted Amplitude: |
DC-to-60 MHz: |
4 divisions. |
DC-to-40 MHz: |
8 divisions. |
HORIZONTAL AMPLIFIER( Input through channel
1 input ) |
X-Y mode: |
switch selectable using X-Y switch. |
CH 1: |
X axis. |
CH 2: |
Y axis. |
Sensitivity: |
1mv/div to 5v/div |
Accuuacy: |
|
Y-Axis: |
± 3%. |
Y-Axis: |
± 6%. |
Input Impedance: |
1MΩ ± 2% //25PF |
Frequency Response: |
DC to 1 MHz typical ( -3 dB ). |
X-Y Phase Difference: |
Approximately 3° at 50 kHz. |
Maximum Input Voltage: |
400v dc+ac pack |
SWEEP SYSTEM |
Sweep Speed: |
0.1 μs/div to 0.5 s/div in 1-2-5 sequence,
21 steps.
Vernier
control porvides fully adjustable sweep time between
steps. |
Accuracy: |
± 3 %. |
Sweep Magnification: |
10X, ± 10% |
Holdoff: |
Continuously adjustable from NORM to 5 times
normal. |
TRIGGERING |
Trigger Modes: |
AUTO ( free run ) or NORM. |
Trigger Source: |
CH1, CH2, ALT, EXT, LINE. |
Maximum External Trigger Voltage: |
300 V dx + AC peak. |
Trigger Coupling: |
AUTO ( AC ): |
30 Hz to 80 MHz. |
TV-H ( AC ): |
Used for triggering from horizontal sync pulses. |
TV-V ( DC ): |
Used for triggering from vertical sync pulses. |
READOUT FUNCTION-Panel setting
desplays |
Vertical axis ( CH1, CH2 ) : |
V/DIV,UNCAL, MAG ( Converted Value) |
Horizontal axis : |
S/DIV, UNCAL, MAG ( Converted Value) |
-Cursor readout function |
Voltage difference △ V : |
△ to REF |
Time difference △ T : |
△ to REF |
Frequency diference 1/ △ T: |
1/1 △ to REF 1 |
NOTE: |
When the X-Y mode, △ V is changed to △ X and △ T to △ Y. |
-Cursor measurement resolution
: |
1/25
DIV |
-Effective cursor range from center
graticule |
Vertical : |
within ± 3 DIV |
Horizontal : |
within ± 4 DIV |
OTHER SPECIFICATIONS |
Calibrating Voltage: |
1 kHz ( nominal ) Positive Square Wave, 0.5
V p-p ( ± 3% ) |
Trace Rotage: |
Electrical, front panel adjustable. |
Power Requirements: |
115/ 230 VAC ± 10%, 50/60Hz , approximately
60 W. |
Dimensions: |
324 x 132 x 398 mm ( 12.8 x 5.2 x 15.6"
). |
Weight: |
Approximately 8.5 kg ( 18.7 1bs ). |